VR06 6 - channel paperless recorder
It is the Low Cost paperless recorder in bigger size 6.4" with highest resolution 640x480 pixels VGA TFT display, infrared detector, maximum 6 channels, plug & play I/O cards. It has many other unique features too,automore dot in com 12 soft keys for easy operation, universal input, the shortest depth 174mm, well designed portable handle, fully isolated channels, reliable and high capacity storage media CF card.
Feature
6.4"Color TFT LCD with 640X480 pixels resolution
The Maximum Channels :
6 isolated analog input channels acim dot in dot th
Plug & Play Supported I / O card, 6 Slots
The High Flexibility :
User configurable l / O card
Expandable modular architecture
Flexible screen configuration
User-Friendly
Soft keys coupled with interactive dialog simplify
setup & operation procedures
Easy - to - access function keys
Infrared Detector:
Shut off LCD automatically to prolong LCD life
and save power while nobody near by
Save Space:
Only 174 mm (6.9") depth behind panel
Various Display Formats:
Vertical trend , Horizontal trend,
Bar Graph, Numerical or mixed
Save Data in Flash ROM,
Compact Flash Card or PC automation dot in dot th
Communication :
Standard Ethernet and optional RS-232 / 422 / 485
The Highest Accuracy
18 - bit A -D analog input, 15-bit D-A analog output.
Fast Sampling Rate:
Within 200 msec for all channels,
Programmable Filter or Moving Average Sampling Method
Statistics with Instant acim dot in dot th, Average , Min./ Max. Values
Progammable Alarms and Messages available
Portable / Bench Top Assembly Kit available<
Specifications
Power
90-250VAC, 47-63Hz, 60VA, 30W maximum
11-18VDC or 18-36 VDC, 60VA, 30W maximum
Display
6.4〞TFT LCD, 640X480 pixel resolution, 256 colors
Memory
8MB storage memory on board
Storage media: 128, 512 MB, 1 GB CF (Compact Flash) cards
Analog Input Cards (AI181, AI182, AI183)
Channels: AI181 ~ 1 channel, AI182 ~ 2 channels, AI183 ~ 3 channels
Resolution: 18 bits thaimeter dot com
Sampling Rate: 5 times/ second
Maximum Rating: -2 VDC minimum, 12 VDC maximum (1 minute for mA input)
Temperature Effect: ±1.5 µV/ µC for all inputs except mA ±3.0 µV/ µC for mA input Sensor Lead Resistance Effect:
Sensor Lead Resistance Effect:
T/C: 0.2 µV/ohm
3-wire RTD: 2.6 µC /ohm of resistance difference of two leads
2-wire RTD: 2.6 µC /ohm of resistance sum of two leads
Burn-out Current: 200nA
Common Mode Rejection Ratio (CMRR): 120dB
Normal Mode Rejection Ratio (NMRR): 55dB
Isolation Breakdown Voltage between channels: 430VAC min.
Sensor Break Detection:
Sensor opened for TC, RTD and mV inputs,
below 1 mA for 4-20mA input,thaimeter dot com below 0.25V for 1-5V inputs,
unavailable for other inputs.
Sensor Break Responding Time:
Within 10 seconds for TC, RTD and mV inputs,
0.1 second for 4-20 mA and 1-5V inputs.
Characteristics:
Type
|
Range
|
Accuracy@ 25 °C
|
Input Impedance
|
J
|
-120 ~ 1000 °C
(-184 ~ 1832 °F)
|
±1 °C
|
2.2MΩ
|
K
|
-200 ~ 1370 °C
(-328 ~ 2498°F)
|
±1 °C
|
2.2MΩ
|
T
|
-250 ~ 400°C
(-418 ~ 752°F)
|
±1 °C
|
2.2MΩ
|
E
|
-100 ~ 900 °C
(-148 ~ 1652 °F)
|
±1°C
|
2.2MΩ
|
B
|
0 ~ 1820 °C
(32 ~ 3308 °F)
|
±2°C
|
2.2MΩ
|
R
|
0 ~ 1768 °C
(32 ~ 3214 °F)
|
±2°C
|
2.2MΩ
|
S
|
0 ~ 1768 °C
(32 ~ 3214 °F)
|
±2°C
|
2.2MΩ
|
N
|
-250 ~ 1300 °C
(-418 ~ 2372 °F)
|
±1 °C
|
2.2MΩ
|
L
|
-200 ~ 900 °C
(-328 ~ 1652 °F)
|
±1 °C
|
2.2MΩ
|
PT100
(DIN)
|
-210 ~ 700 °C
(-346 ~ 1292 °F)
|
±0.4°C
|
1.3KΩ
|
PT100
(JIS)
|
-200 ~ 600 °C
(-328 ~ 1112 °F)
|
±0.4°C
|
1.3KΩ
|
mV
|
-8 ~ 70mV
|
±0.05%
|
2.2MΩ
|
mA
|
-3 ~ 27mA
|
±0.05%
|
70.5Ω
|
0~1V
|
-0.12 ~ 1.15V
|
±0.05%
|
32KΩ
|
0~5V
|
-1.3 ~ 11.5V
|
±0.05%
|
332KΩ
|
1~5V
|
-1.3 ~ 11.5V
|
±0.05%
|
332KΩ
|
0~10V
|
-1.3 ~ 11.5V
|
±0.05%
|
332KΩ
|
Analog Input Cards (AI183V)
Type
|
Range
|
Accuracy@ 25 °C
|
Input Impedance
|
-60~60mV
|
|
±0.1%
|
2.2 MΩ
|
-2~2V
|
-2.2~2.2V
|
±0.3%
|
332KΩ
|
-20~20 V
|
-22~22 V
|
±0.1%
|
332KΩ
|
-20~20 mA
|
-22~22 mA
|
±0.1%
|
70.5Ω
|
Digital Input Card (DI181)
Channels: 6 per card Logic
Low: -5V minimum, 0.8V maximum
Logic High: 2V minimum, 5V maximum
External pull-down Resistance: 1KΩ maximum
External pull-up Resistance: 1.5MΩ minimum
Digital Output Card (DO181)
Channels: 6 per card
Contact Form: N.O. (form A)
Relay Rating: 5A/240 VAC,automore dot com life cycles 200,000 for resistive load
Analog Output Card (AO183I, AO183V)
Channels: 3 per card Output signal: AO183I: 4-20mA, 0-20mA, AO183V: 0-5V, 1-5V, 0-10V
Resolution: 15 bits
Accuracy: ±0.05% of Span ±0.0025% /°C
Load Resistance: 0-500 ohms (current), 10K ohms minimum (voltage)
Output Regulation: 0.01% for full load change thairelay dot com
Output Setting Time: 0.1 second (stable to 99.9%)
Isolation Breakdown Voltage: 1000VAC min.
Integral Linearity Error: ±0.005% of Span
Temperature Effect: ±0.0025% of Span /°C
24VDC Auxiliary Power Supply Card (AP181)
Channels: to be used for 6 transmitters
Output Rating: 24 ± 1 VDC, 180mA in maximum, 30mA / each channel
COMM Module (CM181)
Interface: RS-232 (1 unit), RS-485 or RS-422 (up to 247 units)
Protocol: Modbus Protocol RTU mode
Address: 1-247
Baud Rate: 0.3~38.4 Kbits/sec.
Measured data Bits: 7 or 8 bits
Parity Bit: None, Even or Odd
Stop Bit: 1 or 2 bits automation dot in dot th
Standard Ethernet Communication
Protocol: Modbus TCP/IP, 10 Base T
Ports: AUI (Attachment Unit Interface) and RJ-45, Auto- detect capability
Infrared Detector
Distance: Detect moving human body in distance around 2 meters
Time delayed: 10, 20, 30, 40, 50 or 60 minutes to be defined
Real time clock accuracy vs. temperature inside of housing
Temperature inside housing typical error per month
10 ~ 40 °C 18 seconds
0 °C or 50 °C 52 seconds
-10°C or 60 °C 107 seconds
Environmental & Physical
Operating Temperature: 5 ~ 50 °C
Storage Temperature: -25 ~ 60 °C
Humidity: 20 to 80% RH (non-condensing), maximum relative humidity 80% for temperature up to 31°C automore dot com decreasing linearly to 50% relative humidity at 40°C
Altitude: 2000 M maximum
Insulation Resistance: 20 M ohms min. (at 500 VDC)
Dielectric Strength: 1350 VAC, 50/60 Hz for 1 minute
Vibration Resistance: 10-55 Hz, 10m/ s² for 2 hours
Shock Resistance: 30m/ s² (3g) for operation, 100g for transportation
Operation Position: no inclined restriction
Dimensions: Panel Mount style: 166(W) x 144(H) x 174mm(D)
Bench Top style: 166 (W) x 192 (H) x 194mm (D)
Standard Panel Cutout: DIN size in 138 x 138mm
Approval Standards
Safety: UL61010C-1,
CSA C22.2 No. 24-93
CE: EN61010-1 (IEC1010-1)
over voltage category II, Pollution degree 2
Protective Class:
IP 30 front panel for indoor use,
IP 20 housing and terminals
EMC:
Emission: EN61326 (EN55022 class A, EN61000-3-2, EN61000-3-3)
Immunity: EN61326 (EN61000-4-2, EN61000-4-3, EN61000-4-4,
EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11)
Order Code VR06-1234-567-890
207